EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 137

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
120Operating Manual 1407.6062.32 ─ 08
Note: The workgroup can only be changed after protection level 1 is deactivated (see
chapter 5.2.3.15, "Protection", on page 127).
SCPI command:
:SYSTem:COMMunicate:NETWork[:COMMon]:WORKgroup on page 708
Address Mode
Selects if the IP address is assigned automatically or manually.
"Auto (DHCP)"
The IP address is assigned automatically.
The network used must support automatic assignment of IP address
via DHCP or APIPA (Zeroconf) in order to use this function.
"Static"
The IP address is assigned manually.
SCPI command:
:SYSTem:COMMunicate:NETWork:IPADdress:MODE on page 708
IP Address
Displays the IP address. To enter the IP address manually, select "Static" Address Mode.
SCPI command:
:SYSTem:COMMunicate:NETWork:IPADdress on page 709
Subnet Mask
Displays the Subnet mask. To enter the Subnet mask manually, select "Static" Address
Mode.
This number is used together with the IP address to identify the network segment the
instrument is in.
SCPI command:
:SYSTem:COMMunicate:NETWork[:IPADdress]:SUBNet:MASK on page 709
Default Gateway
Displays the IP address of the default gateway. To enter the default gateway manually,
select "Static" Address Mode.
This address identifies the router on the same network as the instrument that is used to
forward traffic to destinations beyond the local network.
SCPI command:
:SYSTem:COMMunicate:NETWork[:IPADdress]:GATeway on page 709
Restart Network
Shuts down the network connection of the instrument and subsequently re-establishes
the connection.
This function can be used to resolve network problems.
Note: Only the connection of the instrument to the network restarts, the network itself is
not affected.
SCPI command:
n.a.