Instrument Function
R&S
®
SMBV100A
172Operating Manual 1407.6062.32 ─ 08
Target Level
Specifies the nominal level expected at the input of the sensor. The signal generator
adjusts the output power accordingly, in order to meet the target value at the sensor input,
and thus the power required at the DUT.
SCPI command:
[:SOURce<hw>]:POWer:SPC:TARGet on page 682
Limit - RF Level
Sets an upper limit for the RF output power.
You can use it to protect your DUT from damage due to high input power. If you enter an
RF level above this value, the instrument limits the output power to this specified value,
and generates a warning message.
However, the level indication in the status bar is not affected.
Note: The value is not affected by an instrument preset (PRESET key), *RST and the
"Save/Recall" function. It is influenced only by the Factory Preset and the factory value
is equal to maximum level.
SCPI command:
[:SOURce<hw>]:POWer:LIMit[:AMPLitude] on page 678
Catch Range +/-
Sets the capture range of the control system.
Within the range:
Target Level +/- Catch Range
the power control locks and tries to achieve the target level. Readings outside the range
are not considered.
SCPI command:
[:SOURce<hw>]:POWer:SPC:CRANge on page 681
Delay Time
Defines a waiting period between the level adjustment of the generator and the next
measurement of the power sensor.
With this parameter, you can take into account any dead times in the controlled system.
SCPI command:
[:SOURce<hw>]:POWer:SPC:DELay on page 681
Use Peak Power
Activates control by means of the peak power values, provided the power sensor supports
this function. Otherwise, the dialog does not show this parameter.
SCPI command:
[:SOURce<hw>]:POWer:SPC:PEAK on page 682
Use SParameter - Power Sensors
Activates the use of the S-Parameter correction data of the connected power sensor. For
sensors with attenuator this checkbox is automatically checked.
RF Signal and Analog Modulations - RF Block