EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 200

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
183Operating Manual 1407.6062.32 ─ 08
SCPI command:
[:SOURce]:CORRection:CSET:DELete on page 618
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 622
Edit User Cor. Data - User Correction
Calls the editor for editing the selected user correction list.
A list consists of any number of frequency/level value pairs. The currently selected list is
displayed.
Each list is saved as a separate file with extension *.uco. The file name and the directory
to which the file is saved are user-selectable.
Note: Save list only after filling both columns (frequency and level), otherwise the entries
are lost.
"Frequency /
Hz"
Enters the frequency to which the level correction value applies.
Note: The "Fill..." function allows to automatically enter any number of
frequencies with freely selectable range and increment.
Using the "Fill With Sensor" function of the "Edit" sub menu requires
only the entry of the frequency values. The level values are automati-
cally acquired by the connected power sensor.
"Power/dB"
Enters the level correction value to which the specified frequency
applies. The values can be entered manually or automatically with the
"Fill With Sensor" function (available in the "Edit" sub menu).
"Goto"
Selects row for editing.
If Goto row is selected, a window opens for entering the requested row.
RF Signal and Analog Modulations - RF Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals