Instrument Function
R&S
®
SMBV100A
295Operating Manual 1407.6062.32 ─ 08
Parameter Value
Standard GSM
Symbol Rate 270.833 ksymb/s
Coding GSM
Modulation Type MSK
Filter Gauss (FSK)
Filter Parameter BxT 0.3
Power Ramp Control
Attenuation 15 dB
Ramp Time 1 sym
Ramp Function Cosine
Fall Delay 0
Rise Delay 0
Source Internal
State Off
Trigger Mode Auto, Source Internal
Clock
Source Internal
Mode Symbol
SCPI command:
[:SOURce<hw>]:BB:DM:PRESet on page 550
Save/Recall - Digital Modulation
Calls the "Save/Recall" dialog.
From the "Save/Recall" dialog the "File Select" windows for saving and recalling the
complete settings in the "Custom Digital Modulation" menu can be called.
The Digital Modulation Settings are stored as files with the predefined file extension
*.dm. The file name and the directory they are stored in are user-definable.
The complete settings in the "Custom Digital Modulation" dialog are saved and recalled.
SCPI command:
[:SOURce<hw>]:BB:DM:SETTing:CATalog on page 581
[:SOURce<hw>]:BB:DM:SETTing:DELete on page 581
[:SOURce<hw>]:BB:DM:SETTing:LOAD on page 582
[:SOURce<hw>]:BB:DM:SETTing:STORe on page 582
Fast Save
Determines whether the instrument performs an absolute or a differential storing of the
settings.
Baseband Signal - Baseband Block

Table of Contents