Remote Control Commands
R&S
®
SMBV100A
599Operating Manual 1407.6062.32 ─ 08
The numeric suffix to EXTernal<ch> distinguishes between the external trigger via the
TRIGGER 1 (suffix 1) and TRIGGER 2 (suffix 2) connector.
[:SOURce<hw>]:BB:MCCW:TRIGger:ARM:EXECute.........................................................599
[:SOURce<hw>]:BB:MCCW:TRIGger:EXECute.................................................................599
[:SOURce<hw>]:BB:MCCW:TRIGger:EXTernal:SYNChronize:OUTPut...............................600
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:DELay.................................................600
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:INHibit.................................................600
[:SOURce<hw>]:BB:MCCW:TRIGger:RMODe..................................................................601
[:SOURce<hw>]:BB:MCCW:TRIGger:SLENgth.................................................................601
[:SOURce<hw>]:BB:MCCW:TRIGger:SOURce.................................................................602
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal<ch>]:DELay.............................................602
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal<ch>]:INHibit.............................................602
[:SOURce<hw>]:BB:MCCW[:TRIGger]:SEQuence.............................................................603
[:SOURce<hw>]:BB:MCCW:TRIGger:ARM:EXECute
The command stops signal generation for trigger modes Armed_Auto and Armed_Ret-
rigger. A subsequent internal or external trigger event restart signal generation.
Example:
BB:MCCW:TRIG:SOUR INT
sets internal triggering.
BB:MCCW:TRIG:SEQ ARET
sets Armed_Retrigger mode, i.e. every trigger event causes signal
generation to restart.
BB:MCCW:TRIG:EXEC
executes a trigger, signal generation is started.
BB:MCCW:TRIG:ARM:EXEC
signal generation is stopped.
BB:MCCW:TRIG:EXEC
executes a trigger, signal generation is started again.
Usage: Event
[:SOURce<hw>]:BB:MCCW:TRIGger:EXECute
The command executes a trigger. The internal trigger source must be selected using the
command MCCW:TRIGger:SOURce INTernal and a trigger mode other than AUTO
must be selected using the command :BB:MCCW:TRIGger:MODE.
Example:
BB:MCCW:TRIG:SOUR INT
sets internal triggering.
BB:MCCW:TRIG:SEQ RETR
sets Retrigger mode, i.e. every trigger event causes signal gen-
eration to restart.
BB:MCCW:TRIG:EXEC
executes a trigger.
Usage: Event