Example
The following example shows how to set an LF sweep.
1. Set the sweep range.
LFOutput:FREQuency:STARt 4 kHz
LFOutput:FREQuency:STOP 10 kHz
2. Select linear or logarithmic sweep spacing.
LFOutput:SWEep[:FREQuency]:SPACing LIN
3. Set the step width and dwell time.
LFOutput:SWEep[:FREQuency]:STEP[:LINear] 100 Hz
LFOutput:SWEep[:FREQuency]:DWELl 20 ms
4. Determine the sweep mode.
LFOutput:SWEep:MODE AUTO
5. Determine the trigger.
TRIGger0:SOURce SINGle
6. Activate the sweep.
LFOutput:FREQuency:MODE SWEep
7. Trigger the sweep (depending on the mode).
LFOutput:SWEep:EXECute
[:SOURce]:LFOutput:FREQuency....................................................................................648
[:SOURce]:LFOutput:FREQuency:MANual.......................................................................649
[:SOURce]:LFOutput:FREQuency:MODE.........................................................................649
[:SOURce]:LFOutput:FREQuency:STARt.........................................................................650
[:SOURce]:LFOutput:FREQuency:STOP..........................................................................651
[:SOURce]:LFOutput[:STATe]..........................................................................................651
[:SOURce]:LFOutput:SWEep[:FREQuency]:DWELl...........................................................651
[:SOURce]:LFOutput:SWEep[:FREQuency]:EXECute........................................................651
[:SOURce]:LFOutput:SWEep[:FREQuency]:MODE............................................................652
[:SOURce]:LFOutput:SWEep[:FREQuency]:POINts...........................................................652
[:SOURce]:LFOutput:SWEep[:FREQuency]:SHAPe...........................................................653
[:SOURce]:LFOutput:SWEep[:FREQuency]:SPACing........................................................653
[:SOURce]:LFOutput:SWEep[:FREQuency]:STEP[:LINear]................................................654
[:SOURce]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic..........................................654
[:SOURce]:LFOutput<ch>:SHAPe....................................................................................655
[:SOURce]:LFOutput:SIMPedance...................................................................................655
[:SOURce]:LFOutput:VOLTage........................................................................................655
[:SOURce]:LFOutput:FREQuency <Frequency>
The command sets the frequency of the LF signal for CW mode (:SOUR:MODE CW mode).
The setting is valid for all analog modulations (AM/FM/PhiM/Pulse) with internal modu-
lation source.