EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 667

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100A
650Operating Manual 1407.6062.32 ─ 08
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
The instrument operates in fixed-frequency mode. CW and
FIXed are synonyms.
The output frequency is set with [:SOURce]:LFOutput:
FREQuency.
SWEep
The instrument operates in SWEep mode. The frequency is set
using the commands [:SOURce]:LFOutput:FREQuency:
STARt, [:SOURce]:LFOutput:FREQuency:STOP or [:
SOURce]:LFOutput:FREQuency:MANual.
*RST:
CW
Example:
LFO:FREQ:MODE SWE
sets the SWEep mode. The settings under
SOURce:LFOutput:FREQuency:STARt; STOP; MANual
become effective.
[:SOURce]:LFOutput:FREQuency:STARt <Start>
This command sets the start frequency for the LF Sweep mode.
Parameters:
<Start> float
Range: 0.1 Hz to 1 MHz
Increment: 0.1 Hz
*RST: 1 kHz
Example:
RST*
activates all presettings.
LFO:SWE:MODE AUTO
sets the AUTO sweep mode, i.e. each trigger triggers a complete
sweep.
TRIG0:SOUR SING
sets the SINGle trigger mode, i.e. the sweep is triggered by the
command :LFOutput:SWEep:EXECute or *TRG.
LFO:FREQ:STAR 100 kHz
sets the start frequency for the LF sweep to 100 kHz.
LFO:FREQ:STOP 200 kHz
sets the stop frequency of the LF sweep to 200 kHz.
LFO:FREQ:MODE SWE
sets the LF sweep mode.
LFO:SWE:EXEC
a one-off LF sweep from 100 kHz to 200 kHz is performed. The
linear step width is 1 kHz with a dwell time of 15 ms (preset values).
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals