Sirona Dental Systems GmbH 4Operation
Operating Instructions INTEGO 4.5Dentist element
båÖäáëÜ
10. Attach the required root canal file to the handpiece.
11. Short the electrical measurement system. Plug in the files and hold
them directly against the mucosal electrode. This cancels out any
inaccuracies caused by jumps in impedance in the measurement
setup (standardization).
If a short signal is heard and the distance display appears with no
bars, the standardization was successful. If not, check the
electrical connections for signs of damage.
12. Place the mucosal electrode in the patient's mouth and perform the
treatment. Activate the bur using the foot pedal.
The measured root canal depth is displayed by a colored bar in
the distance display. For more information, please refer to
"Distance display" [ 123].
CAUTION
Preventing incorrect measurements
When performing the apex measurement, wear non-conductive gloves
to avoid measurement errors caused by unwanted leakage current.
During the measurement, the instrument must not come into contact
with the patient's mucosa, metallic tooth restorations, or the mucosal
electrode. It is recommended to pull the silicone isolation sleeve over
the handpiece and perform the treatment using a cofferdam.