4Operation Sirona Dental Systems GmbH
4.7Water unit Operating Instructions INTEGO
4. Confirm your selection by pressing the favorite key
2 / Setup
.
Switching starts. A revolving element is shown on the EasyPad
display. If there is still disinfectant in the tank, it is rinsed down the
drain.
The
Refill water
display appears.
5. Mix distilled water with the DENTOSEPT P disinfectant at a ratio of
100:1 (1 liter of water, 10 ml DENTOSEPT P) and fill this into the
disinfectant tank of the water unit. The tank has a capacity of approx.
1.3 liters. It is full when the water surface is visible on the filter of the
filling funnel.
The revolving element is displayed again on the EasyPad display.
6. Wait until switching the mode of operation is complete.
The adjacent message "Aqua" is shown. The disinfection system
is now changed to stand-alone water supply.
7. Acknowledge the message by pressing the favorite key
2 / Setup
.
The time is displayed.
Resetting disinfection system for operation with public drinking water
and disinfectants
Resetting the disinfection system to public drinking water is similar to
changing to stand-alone water supply. Instead of the
Refill water
display,
the
Refill disinfectant
display appears on the EasyPad.
1. Fill approx. 0.5 liters of DENTOSEPT P into the disinfectant tank.
2. Sanitize the water lines after switching to the public drinking water
supply, see “Sanitizing with disinfection system” [ 277].
Level control
If the
Refill water
or
Refill disinfectant
display appears during the
treatment, the reservoir is almost empty (< 300 ml). Pressing the
Counterclockwise rotation/user profile
key hides the display and the
treatment is continued. Refill the tank as soon as possible.
Water unit, disinfectant tank < 300 ml
2
Setup
1LH O +
10 mLDesinf.
2
DENTOSEPT PDENTOSEPT P
2
Setup
A/B
IMPORTANT
300 ml are approximately sufficient for:
➢ 5 min. spray operation (motor, turbine, SIROSONIC L) or
➢ 2 min. SPRAYVIT operation or
➢ 4 x tumbler fillings
The actual values depend on the instrument types and settings used.