Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.3Instruments and instrument hoses
båÖäáëÜ
● Leave the instruments in the water unit
The autopurge function remains activated.
The instruments remain in the Comfort water unit and the treatment
center can be switched off at the mains. On the next day, the
autopurge function is automatically performed again on all of the
instruments remaining in the water unit as well as on the tumbler
filling unit immediately after the treatment center is switched on.
Then you can prepare the treatment center for daily practice
operation.
If the treatment center is out of operation over a prolonged period of
time, you can briefly switch it on every day and then switch it off again
when the purging process is finished. This ensures that the number
of microorganisms in the water lines will not increase excessively.
Empty the rinsing tumbler after each purging process and place the
empty tumbler again below the tumbler outlet.
If any instruments are removed from the water unit or new
instruments are connected to the treatment center while the latter is
switched off, they must be plugged into the adapters of the water unit
or returned to the instrument holder again before switching the
treatment center back on.
● Deposit instruments
The autopurge function is completed.
➢ Remove the instruments from the Comfort water unit and put them
back in place.
The autopurge procedure is finished. The treatment center is again
ready for operation and can be prepared for the practice day.
Canceling the autopurge function
In the case of the error message
Deposit instruments
or during purging,
the autopurge function can be canceled.
➢ Press the
Counterclockwise rotation/user profile
fixed key.