5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.3Instruments and instrument hoses Operating Instructions INTEGO
3. Dentist element side: Insert the couplings of all water-carrying
treatment instruments into the adapter in the Comfort water unit (for
SPRAYVIT E: valve lever in the up position, locking knob in the down
position).
Also for the CS dentist element: Move the swivel arm of the
instruments to be purged into the operating position so that the weight
of the instrument hoses allows them to remain in this position.
4. Assistant element side: Insert the valve bodies of the SPRAYVIT E
into the adapter on the water unit.
All water-carrying instruments are inserted into the cleaning
adapters.
5. Press the favorite key
2 / Setup
.
Automatic purging of the water lines will start. A revolving
element is shown next to the
AutoPurge
display on the EasyPad.
Purging water lines
The removed instruments are purged with water for the duration of the set
purge time. Then the tumbler filler is purged.
Following the purging process, the
End
display appears.
You can now continue in one of two ways:
IMPORTANT
Pinching the instrument hoses
Be careful not to pinch the instrument hoses when inserting the
instruments.
With treatment centers with the Ambidextrous option: convert the water
unit to the left side so that the adapters on the water unit point towards
the patient chair, see "Converting the Turn water unit for right-handers /
light-handers" [ → 161].
If the hoses are pinched, the water flow will be obstructed during
purging.
The water flow rate to the instruments is not checked by the treatment
center.