5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.3Instruments and instrument hoses Operating Instructions INTEGO
All water-carrying instruments are inserted into the cleaning
adapters.
5. Touch the
Start
key on the touchscreen.
The purging process begins.
Purging water lines
The removed instruments are purged with water for the duration of the set
purge time. Then the tumbler filler is purged. The progress bar displayed
on the touchscreen refers to the entire autopurge program, not to the set
instrument purge time.
You can now continue in one of two ways:
● Leave the instruments in the water unit
The autopurge function remains activated.
The instruments remain in the water unit and the treatment center can
be switched off at the mains. On the next day, the autopurge function
is automatically performed again on all of the instruments remaining
in the water unit as well as on the tumbler filling unit immediately after
the treatment center is switched on.
Then you can prepare the treatment center for daily practice
operation.
If the treatment center is out of operation over a prolonged period of
time, you can briefly switch it on every day and then switch it off again
when the purging process is finished. This ensures that the number
of microorganisms in the water lines will not increase excessively.
Empty the rinsing tumbler after each purging process and place the
empty tumbler again below the tumbler outlet.
If any instruments are removed from the water unit or new
instruments are connected to the treatment center while the latter is
switched off, they must be plugged into the adapters of the water unit
or returned to the instrument holder again before switching the
treatment center back on.
● Deposit instruments
The autopurge function is completed.
IMPORTANT
Pinching the instrument hoses
Be careful not to pinch the instrument hoses when inserting the
instruments.
With treatment centers with the Ambidextrous option: convert the water
unit to the left side so that the adapters on the water unit point towards
the patient chair, see "Converting the Turn water unit for right-handers /
light-handers" [ → 161].
If the hoses are pinched, the water flow will be obstructed during
purging.
The water flow rate to the instruments is not checked by the treatment
center.