5.6Sanitizing Operating Instructions INTEGO
Allow 24 hours reaction time
In order to effectively combat germs, you must allow the DENTOSEPT P
disinfectant to act at least 24 hours, but for no longer than 3 days
(maximum sanitation time).
1. Switch the treatment center off via the mains switch A on the base of
the chair. Make sure that the treatment center is switched off for at
least 24 hours, but not longer than 3 days (maximum sanitation time).
2. Empty the rinsing tumbler and return the empty tumbler below the
tumbler outlet.
➢ Switch the treatment center on again after 24 hours and before 3
days have elapsed.
Filling the fresh water bottle with water
✔ The display
Refill water
appears on the EasyPad.
1. Unscrew the fresh water bottle counterclockwise from the holder on
the water unit and empty the fresh water bottle completely.
2. Mix food-grade water (< 100 CFU/ml) with the DENTOSEPT P
disinfectant at a ratio of 100:1 (1 liter of water, 10 ml DENTOSEPT P)
and fill this into the fresh water bottle. The bottle has a capacity of
approx. 1.3 liters.
3. Attach the hose to the fresh water bottle and screw the bottle
clockwise into the holder.
4. Press the favorite key
2 / Setup
.
A revolving element is shown on the EasyPad display next to the
text "AqU". Tumbler filling is switched on and is flushed with
water.
Purging water lines with water
The DENTOSEPT P disinfectant is rinsed out of the instrument hoses,
the SPRAYVIT hoses, and tumbler filler with water.
✔ The tumbler filling was automatically flushed with water.
1. Hold the 3-way syringe over the cuspidor or over a watertight
container with sufficient capacity and press the water key for at least
15 seconds.
2. Remove a single instrument from holder and hold it over the cuspidor
or over a watertight container with sufficient capacity. For the CS
dentist element, move the swivel arm of the instrument into the
working position.
3. Step on the pedal of the foot control briefly.
The removed instrument is purged with water for approx. 15
seconds. The purging process stops automatically.