Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.6Sanitizing
båÖäáëÜ
4. Afterward, return the instrument to the instrument holder.
5. Repeat the procedure for all other instruments.
6. After you have flushed the last instrument, press favorite key
2 /
Setup
.
The time is displayed on the EasyPad display.
7. Empty the rinsing tumbler and return the empty tumbler below the
tumbler outlet.
The sanitation process is finished. The treatment center is again
ready for operation.
Interrupting instrument purging
➢ Step on the foot pedal to interrupt the instrument purging process. If
the instrument is not put back in the holder, the purging process can
be continued by pressing the foot pedal again.
Canceling the sanitizing process
The sanitizing process can be canceled, e.g., if the treatment center must
be switched on again before 24 hours have passed.
✔ The sanitizing process has started.
➢ Press the
Counterclockwise rotation/user profile
key.
The sanitation process is canceled. If the water lines have
already been treated with DENTOSEPT P disinfectant, they have
to be rinsed with water before use. The display
Refill water
appears, see "Filling the fresh water bottle with water" (above).
Cleaning the suction hoses
After sanitizing with the fresh water bottle, the suction hoses should also
be chemically cleaned, see "Cleaning suction system using cleaning
adapter in the cuspidor or via external container" [ → 242].