Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.6Sanitizing
båÖäáëÜ
Treating water lines with DENTOSEPT P disinfectant
The water lines of the instrument hoses, SPRAYVIT hoses, and the
tumbler filler are filled with pure DENTOSEPT P.
Tumbler filling, discolorations due to DENTOSEPT P
Allow 24 hours reaction time
In order to effectively combat germs, you must allow the DENTOSEPT P
disinfectant to act at least 24 hours, but for no longer than 3 days
(maximum sanitation time).
1. Switch the treatment center off via the mains switch A on the base of
the chair. Make sure that the treatment center is switched off for at
least 24 hours, but not longer than 3 days (maximum sanitation time).
2. Empty the rinsing tumbler and return the empty tumbler below the
tumbler outlet.
Switch the treatment center on again after 24 hours and before 3
days have elapsed.
Error message: Inserting instruments in the water unit
If an instrument is accidentally deposited in its holder during the reaction
time, the message
Insert instruments into water unit
is displayed on the
EasyPad display after the treatment center is switched on. The sanitizing
process is then completed anyway.
1. Place the accidentally removed instrument back in the adapter on the
water unit so that it can be flushed.
2. If you want to exclude the concerned instruments from the sanitation
process, press favorit key
2 / Setup
.
The instrument marked on the EasyPad display will not be
flushed.
Stand-alone water supply only: refilling water
When operating with a stand-alone water supply,
Refill water
is displayed.
Mix distilled water with the DENTOSEPT P disinfectant at a ratio of
100:1 (1 liter of water, 10 ml DENTOSEPT P) and fill this into the
disinfectant tank of the water unit.
NOTICE
Splashes of pure DENTOSEPT P can result in discoloration of surfaces
if left for too long.
Any splashes should therefore be removed immediately using a moist
cloth.
A
IMPORTANT
Blocking water and air supply
When the treatment center is turned on again after 24 hours, the
sanitizing process continues automatically. However, if the water and air
supplies are switched off, the DENTOSEPT P disinfectant cannot be
rinsed out of the water lines.
Switch the treatment center back on after 24 hours only when the water
and air supply lines are open.