5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.6Sanitizing Operating Instructions INTEGO
When sufficient water has been refilled, the sanitizing program
will continue.
Purging water lines with water
The DENTOSEPT P disinfectant is rinsed out of the SPRAYVIT hoses,
instrument hoses, and tumbler filling unit with water. This takes several
minutes.
Returning instruments to their holders
After the water lines have been rinsed, the instruments can be removed
from the cleaning adapters on the water unit and returned to their holders.
The text
End
is displayed on the EasyPad display.
1. Place the SPRAYVIT sleeves and the straight and contra-angle
pieces on the instruments. Place all the instruments back in their
holders.
The time is displayed on the EasyPad display.
2. Empty the rinsing tumbler and return the empty tumbler below the
tumbler outlet.
The sanitation process is finished. The treatment center is again
ready for operation.
Canceling the sanitizing process
The sanitizing process can be canceled, e.g., if the treatment center must
be switched on again before 24 hours have passed.
The sanitizing process has started.
Press the
Counterclockwise rotation/user profile
key.
The sanitation process is canceled. If the water lines have
already been treated with DENTOSEPT P disinfectant, they have
to be rinsed with water before use. If
Refill water
appears during
operation with a stand-alone water supply, see "Stand-alone
water supply only: refilling water" (above).
Cleaning the suction hoses
After sanitization, all suction hoses should also be chemically cleaned:
If the treatment center is not equipped with the chemical suction hose
cleaning option, see "Cleaning suction system using cleaning
adapters in the cuspidor or an external container" [ 242].
If the treatment center is equipped with the chemical suction hose
cleaning option, see "Suction hose cleaning on the Comfort water
unit" [ 237].