5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.6Sanitizing Operating Instructions INTEGO
All instruments which can be sanitized are inserted into the
cleaning adapters.
5. Touch the
Start
key on the touchscreen.
Sanitation phase 2 – Checking the water flow
First, the treatment center checks whether there is water flow through the
instruments.
Sanitization phase 2 is highlighted on the touchscreen.
Wait briefly until the water flow has been checked.
If sufficient water flow is present, the treatment center continues
with sanitation phase 3.
Error message: No water flow
If the treatment center detects no water flow through an instrument or
through the tumbler filler, you can try to restore water flow. If this is not
possible, the instrument concerned can be excluded from the sanitation
process.
1. Check the water flow through the instruments in the instrument
positions marked with a warning triangle on the touchscreen. Set the
instruments to maximum water flow. Leave all instruments plugged
into the water unit.
If the treatment center detects the water flow, the warning triangle
will disappear. If there is sufficient water flow through all
instruments, sanitation phase 3 automatically continues.
2. If you want to exclude the instruments concerned from sanitation,
touch the
Exclude instrument
key.
The treatment center continues with sanitation phase 3. The
water lines you excluded are not included in the sanitation
process.
If it is impossible to restore the flow to the tumbler filling unit, sanitation is
not possible. The tumbler filling unit cannot be excluded from the
sanitation process.
IMPORTANT
Pinching the instrument hoses
Be careful not to pinch the instrument hoses when inserting the
instruments.
With treatment centers with the Ambidextrous option: convert the water
unit to the left side so that the adapters on the water unit point towards
the patient chair, see "Converting the Turn water unit for right-handers /
light-handers" [ 161].
If the hoses are pinched, the water flow will be obstructed during
purging.
Start
Start
1
2
3
4
5
6
H
2
O
Desinf
Stop
!