Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.6Sanitizing
båÖäáëÜ
Sanitation phase 3 – Treating water lines with DENTOSEPT P
disinfectant
The water is pumped out of the water tank of the water unit via the tumbler
filler. Then the water tank is automatically filled with pure DENTOSEPT
P. This activates flushing of the instrument hoses, the SPRAYVIT hoses
and the tumbler filler with the DENTOSEPT P disinfectant.
Tumbler filling, discolorations due to DENTOSEPT P
Sanitation phase 4 – Allow 24 hours reaction time
In order to effectively combat germs, you must allow the DENTOSEPT P
disinfectant to act at least 24 hours, but for no longer than 3 days
(maximum sanitation time).
✔ Sanitization phase 4 is highlighted on the touchscreen.
✔ The treatment center displays the remaining reaction time on the
touchscreen next to the sanitation phase 4 field, starting from 24
hours.
✔ The treatment center has automatically switched to Standby mode.
1. Leave all instruments plugged into the water unit.
2. Switch the treatment center off via the mains switch A on the base of
the chair. Make sure that the treatment center is switched off for at
least 24 hours, but not longer than 3 days (maximum sanitation time).
3. Empty the rinsing tumbler and return the empty tumbler below the
tumbler outlet.
4. Switch the treatment center on again after 24 hours and before 3
days have elapsed.
NOTICE
Splashes of pure DENTOSEPT P can result in discoloration of surfaces
if left for too long.
Any splashes should therefore be removed immediately using a moist
cloth.
IMPORTANT
Blocking water and air supply
If the treatment center is turned on again after 24 hours, the sanitation
process continues automatically with sanitation phase 5. However, if the
water and air supplies are switched off, the DENTOSEPT P disinfectant
cannot be rinsed out of the water lines.
Switch the treatment center back on after 24 hours only when the water
and air supply lines are open.