5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.6Sanitizing Operating Instructions INTEGO
Error message: Inserting instruments in the water unit
If an instrument is accidentally deposited in its holder during the reaction
time, the message
Insert instruments into water unit
is displayed on the
touchscreen after the treatment center is switched on. The sanitizing
process is then completed anyway.
1. Remove the accidentally removed instrument and place it back in the
adapter on the water unit so that it can be rinsed in sanitation phase 5.
2. If you want to exclude the instruments from sanitation, touch the
Exclude instrument
key.
Excluded instruments will not be rinsed in sanitation phase 5.
Stand-alone water supply only: refilling water
When operating with a stand-alone water supply,
Refill water
is displayed.
Mix distilled water with the DENTOSEPT P disinfectant at a ratio of
100:1 (1 liter of water, 10 ml DENTOSEPT P) and fill this into the
disinfectant tank of the water unit.
When sufficient water has been refilled, the sanitizing program
will continue.
Sanitation phase 5 – Rinsing water lines with water
The DENTOSEPT P disinfectant is rinsed out of the SPRAYVIT hoses,
instrument hoses, and tumbler filling unit with water. This takes several
minutes.
Sanitation phase 6 – Returning instruments to their holders
After the water lines have been rinsed, the instruments can be removed
from the cleaning adapters on the water unit and returned to their holders.
The
Return instruments to their holders
display appears on the
touchscreen.
1. Place the SPRAYVIT sleeves and the straight and contra-angle
pieces on the instruments. Place all the instruments back in their
holders.
2. Empty the rinsing tumbler and return the empty tumbler below the
tumbler outlet.
The sanitation process is finished. The treatment center is again
ready for operation.