EasyManuals Logo

Topcon 3D OCT-1 User Manual

Topcon 3D OCT-1
236 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
121
DISPLAYING TOMOGRAMS
3.9. Pin-point Registration™
3.9.1. Shadowgram
A shadowgram image is created by summing the 50 pixels under RPE
longitudinally on the 3D tomogram data, and is displayed as a B/W
image (see right). The fundus features (for example, blood vessel pat-
tern) are visible in the shadowgram image.
This software performs "Pin-point Registration™" by using the shadow-
gram data.
3.9.2. Automatic Positioning
After analysis, the shadowgram image and color fundus image are positioned automatically.
3.9.3. Manual Positioning
To Perform a Manual Positioning:
1 Click Overlay Reposition.
2 Place the cursor on the image. While depressing the left mouse button, drag the image to the
desired location.
3 The shadowgram image will become transparent to allow the user to move and align it to the
appropriate position of the color fundus image.
CAUTION
As the result of automatic positioning, the fundus image is sometimes
deformed to the oval shape on the fundus/anterior segment image dis-
play area (A-3). This is the result of the positioning process and there is
no problem.
CAUTION
After analysis, the images are automatically positioned. If sufficient data
required for overlapping are not provided, automatic positioning may not
be done correctly.
Sample projection image

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon 3D OCT-1 and is the answer not in the manual?

Topcon 3D OCT-1 Specifications

General IconGeneral
BrandTopcon
Model3D OCT-1
CategoryMedical Equipment
LanguageEnglish

Related product manuals