<5. Functions>
143
IM 01E21A02-03EN
5.10.4 Test Mode Auto Reset
Ifthespeciedtimelapseswithnoparameterschangedwhilethetestmodeisenabled,thetest
mode is reset automatically. When any test mode parameter is changed, the test mode reset time
is extended.
Thesettingcanbeconguredwiththefollowingparameter.
Menu path
Display Devicesetup►Diag/Service►Test►(seebelow)
F
OUNDATION
Fieldbus
DeviceConguration►MTB►DeviceConguration►QueryDevice►MaintTBOriginal
Parameters(Part2)►Page4►Test
Block
Name
Relative
Index
Parameter
Description
Display F
OUNDATION Fieldbus
MTB 95 Release time Test Auto Release Time
Speciesthetimeto
automatically reset the test
mode.
From the table below, select the test mode auto reset time.
Selection
Description
Display
F
OUNDATION
Fieldbus
10 min 10min Sets the reset time to 10 min.
30 min 30min Sets the reset time to 30 min.
1 h 1h Sets the reset time to 1 hour.
3 h 3h Sets the reset time to 3 hours.
12 h 12h Sets the reset time to 12 hours.

Table of Contents

Other manuals for YOKOGAWA ADMAG TI Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG TI Series and is the answer not in the manual?

YOKOGAWA ADMAG TI Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG TI Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals