3.4 Integration of DD
IfthehostsupportsDD(DeviceDescription),theDDofthisproductneedstobespecied.Check
if the host has the following directory under the directory to specify DD.
594543¥0016(AXG)
594543¥0017(AXW)
 (594543isthemanufacturernumberofYokogawaElectricCorporation,and0016is
thedevicenumberofAXG,and0017isthedevicetypeforAXW.)
If the directory is not found, the DD of this product has not been included. Create the above
directoryandcopytheDDleofthisproduct(0m0n.󰀨5,0m0n.sy5(m,nisanumeral))
(separatelysupplied)intothedirectory.
Here,’0m’inthelenameshowsthedevicerevision,and’0n’showstheDDrevision.
Once the DD is installed in the directory, the name and attribute of all parameters of this product
are displayed.
Also,o󰀨-linecongurationispossiblebyusingcapabilitiesles(CFF).
IfyoudonothavetheDDorcapabilitiesles,youcandownloadthemfromYokogawa’sweb
page: Access the following web site, or contact your sales agent where you purchased this
product.
http://www.yokogawa.com/d/doc/
3.5 Reading the Parameters
Select the AI block of this product from the host screen, and read the OUT parameter. The
measuredvalueassignedtothepresentAIisdisplayed.CheckthatMODE_BLKofthefunction
block and resource block is set to AUTO, change the input signal being measured, and read the
parameter again. A new designated value will be displayed.
3.6 Continuous Record of Values
If the host has a function that continuously records the indications, use the function to record the
indications(values).Dependingonthehostbeingused,itisnecessarytosetthescheduleof
Publish(functiontotransmittheindicationtothebusonaperiodicbasis).
3.7 Generation of Alarm
Theblockalarm,out-of-rangealarmoftheoutputparameter(OUT)andupdatealarmatsetting
change can be generated from this product. When generating alarm, a Link Object and a VCR
Static Entry need to be set. For details of Link Object and VCR Static Entry, refer to Subsection
4.6.1 Link object and Subsection 4.5.1 VCR Setting.

Table of Contents

Other manuals for YOKOGAWA ADMAG TI Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG TI Series and is the answer not in the manual?

YOKOGAWA ADMAG TI Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG TI Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals