5.9.6 Verication(DeviceHealthDiagnosis)Function
Thevericationfunctiondiagnosesthehealthoftheproductanddisplaysthediagnosisresult.It
takesabout15minutestocompletethevericationfunction,andtheprogressoftheverication
function can be checked via the bar graph and the remaining time shown on the display.
10:00
Verification execution
Excitation circuit...
01 : 00
SET SFT INC
This function can inspect each condition of the magnetic circuit, excitation circuit, and calculation
circuit, and executes the device health diagnosis of the product based on their internal alarm
record along with their wiring misconnection. This function evaluates the diagnosis result based
ontheconditionsoftheproductandshows“Passed”ifnoproblemisfoundor“Failed”ifa
problem is found on the display.
The following execution results are shown on the display.
• Examples of “Passed” results
VF check results Passed
VF Operation time 00001D 10:01
Magnetic circuit result Passed
Exciting circuit result Passed
Calculation circuit result Passed
Device status result Passed
Connection status result Passed
• Examples of “Failed” results
VF check results Failed
VF Operation time 00001D 10:01
Magnetic circuit result Passed
Exciting circuit result Passed
Calculation circuit result Failed
Device status result Passed
Connection status result Passed
ContactYokogawaservicecenterifthe“Failed”messageisdisplayedforitemsfrom“Magnetic
circuitresult”to“Devicestatusresult”.
If“Failed”isdisplayedfor“Connectionstatusresult”,checkforwiringmisconnectionbetweenthe
sensorandtransmitter(refertoSubsection5.9.5fordetail)anddamageonthecables.
Thevericationfunctioncanbeusedintwowaysdependingonthestateoftheuid;onestate
whereauidisowingandanotherstatewherenouidisowing.
Thetwodiagnosisresults(currentandprevious)arestoredinthedevicememory,andtheycan
be checked later.
Ifthevericationfunctionisused,thefollowingresultscanbedisplayed.
Total judgment result Calculation circuit diagnosis result
Operationtimeofvericationfunction Device alarm diagnosis result
Magnetic circuit diagnosis result Wiring connection diagnosis result
Excitation circuit diagnosis result
Thesettingcanbeconguredwiththefollowingparameters.
Menu path
Display Devicesetup►Diag/Service►Diagnosis►(seebelow)
F
OUNDATION
Fieldbus
Diagnostic►DTB►DeviceDiagnostics►Diagnostics/Alerts►Diagnostics►(see
below)