<5. Functions>
62
IM 01E21A02-03EN
5.1.9 Density Setting
Thedensitymustbesetinordertomeasurethemassowrate.Thedensitycanbeselected
fromthexeddensityordensitycorrectedwithtemperature.
Fordetailsabouthowtomeasurethemassowrateorcorrectthedensitybytemperature,refer
to Subsection 5.5.5.
Ifdensityissetto“0”whilethemassowrateismappedtoPV,asettingerrorwillresult.
Thesettingcanbeconguredwiththefollowingparameters.
Menu path
Display Devicesetup►Detailedsetup►Provar►Density►(seebelow)
F
OUNDATION
Fieldbus
DeviceConguration►STB►DeviceConguration►Conguration►Process
Variables►Density►(seebelow)
Block
Name
Relative
Index
Parameter
Description
Display F
OUNDATION Fieldbus
STB 97 Value select Density Select
Selects one of the following
temperature-based density
corrections.
Fixed value : Uses the
xeddensity
Correction value : Uses the
corrected density
(OnlyavailableforAXG,notfor
AXW).
STB 98 Unit Density Unit Speciestheunitofthedensity.
STB 99 Fixed density Fixed Density
Speciesthevalueofthexed
density.
STB 100 Std density Standard Density
Speciesthereference
standard density value to
use the temperature-based
density correction function
(onlyavailableforAXG,notfor
AXW).
STB 105 Correct density CORRCT Density
Displays the corrected density
(onlyavailableforAXG,notfor
AXW).