<5. Functions>
62
IM 01E21A02-03EN
5.1.9 Density Setting
Thedensitymustbesetinordertomeasurethemassowrate.Thedensitycanbeselected
fromthexeddensityordensitycorrectedwithtemperature.
Fordetailsabouthowtomeasurethemassowrateorcorrectthedensitybytemperature,refer
to Subsection 5.5.5.
Ifdensityissetto“0”whilethemassowrateismappedtoPV,asettingerrorwillresult.
Thesettingcanbeconguredwiththefollowingparameters.
Menu path
Display Devicesetup►Detailedsetup►Provar►Density►(seebelow)
F
OUNDATION
Fieldbus
DeviceConguration►STB►DeviceConguration►Conguration►Process
Variables►Density►(seebelow)
Block
Name
Relative
Index
Parameter
Description
Display F
OUNDATION Fieldbus
STB 97 Value select Density Select
Selects one of the following
temperature-based density
corrections.
• Fixed value : Uses the
xeddensity
• Correction value : Uses the
corrected density
(OnlyavailableforAXG,notfor
AXW).
STB 98 Unit Density Unit Speciestheunitofthedensity.
STB 99 Fixed density Fixed Density
Speciesthevalueofthexed
density.
STB 100 Std density Standard Density
Speciesthereference
standard density value to
use the temperature-based
density correction function
(onlyavailableforAXG,notfor
AXW).
STB 105 Correct density CORRCT Density
Displays the corrected density
(onlyavailableforAXG,notfor
AXW).