<5. Functions>
88
IM 01E21A02-03EN
5.5.6 Calorie Calculation
Byobtainingtheabsolutetemperatureorthetemperaturedi󰀨erencefromotherproduct
(temperaturetransmitter,etc.)fromMAO,thecaloriecalculationisexecutable.
The calorie is obtained using the following formula.
Q=c×V
m
×∆t
Q:Calorie[J/s]
c:Specicheat[J/kg•K]
V
m
:Massowrate[kg/s]
∆t:Temperaturedi󰀨erence[K]
Thetemperaturedi󰀨erenceofΔtvariesdependingontheMAOfunctionsetinSection5.4.Ifthe
temperaturedi󰀨erenceisselected,thetemperatureobtainedinMAOisused.Iftheabsolute
temperatureisselected,adi󰀨erencebetweenthetemperatureobtainedinMAOandthe
reference temperature set in Subsection 5.1.10 is used.
[“TemperatureacquiredbyMAO”–“Setreferencetemperature”]
ThisfunctioncanbeusedonlyforAXG,notforAXW.Thesettingcanbeconguredwiththe
following parameter.
Menu path
Display Devicesetup►Detailedsetup►Provar►Calorie►(seebelow)
F
OUNDATION
Fieldbus
DeviceConguration►STB►DeviceConguration►Conguration►Process
Variables►Calorie►(seebelow)
Block
Name
Relative
Index
Parameter
Description
Display F
OUNDATION Fieldbus
STB 106 Specicheat SpecicHeat Speciesthespecicheat.
Thecaloriecalculationcanbeconguredusingthefollowingprocedures.
• When the temperature is selected as the external temperature input function:
(1) Setthefunctionoftheexternaltemperatureinputfunctiontothetemperaturebyreferringto
Section 5.4.
(2) SettherangehighlimitandlowlimitoftheexternaltemperatureinputbyreferringtoSection
5.4.
(3) SetthereferencetemperaturebyreferringtoSubsection5.1.10.
(4) Setthespecicheat.
•Whenthetemperaturedi󰀨erenceisselectedastheexternaltemperatureinput
function:
(1) Settheexternaltemperatureinputfunctiontothetemperaturedi󰀨erencebyreferringto
Section 5.4.
(2) SettherangehighlimitandlowlimitoftheexternaltemperatureinputbyreferringtoSection
5.4.
(3) Setthespecicheat.

Table of Contents

Other manuals for YOKOGAWA ADMAG TI Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG TI Series and is the answer not in the manual?

YOKOGAWA ADMAG TI Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG TI Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals