<5. Functions>
61
IM 01E21A02-03EN
Whenaowrateisdecreasingandanactualowrategoesbelow0.95[m
3
/h],theoutputow
rateisreducedto0.0[m
3
/h]withthelow-cutfunction.Ontheotherhand,whenaowrateis
increasingandtheactualowrategoesover1.05[m
3
/h],aowrateisoutput.
F0401.ai
0
Output [m
3
/h]
1
Hysteresis
1 %
Input [m
3
/h]
1
(1)
(2)
1.050.95
NOTE
Notethatthetotalizationmightbecountedduetotheinuenceoftheoutputuctuationnear0%
output if a small low-cut value is set.
Inparticular,ifthevalueoftheowratespan,dampingtimeconstantorconductivityislow,the
totalizationiseasilycountedwhentheowrateiszero.Insuchacase,increasetheowrate
span, damping time constant, or low-cut value.
NOTE
If the output process value to be output is changed, it is necessary to specify the low-cut value
again.
5.1.8 Sensor’s Nominal Size Setting
To combine the remote transmitter with other remote sensor, the nominal size of the remote
sensor must be set.
Thesettingcanbeconguredwiththefollowingparameters.
Menu path
Display Devicesetup►Detailedsetup►Sensor►(seebelow)
F
OUNDATION
Fieldbus
DeviceConguration►STB►DeviceConguration►Conguration►Sensor►(see
below)
Block
Name
Relative
Index
Parameter
Description
Display F
OUNDATION Fieldbus
STB 57 Nominal size unit Nominal Size Unit
Speciestheunitofthe
nominal size.
STB 58 Nominal size Nominal Size Speciesthenominalsize.
NOTE
For the integral type, do not change the nominal size and its unit since they are set at factory
shipment.