<5. Functions>
59
IM 01E21A02-03EN
NOTE
Theoutputuctuationincreasesifthedampingtimeconstantissettoalowervalue.
Set the damping time constant to 5 seconds or longer for using the constant for control purpose.
5.1.7 Low-cut Function Setting
A low-cut value can be set for the frequency output, pulse output and totalizer.
Ifthelow-cutfunctionisused,theowratebelowsetpontscanbestoppedfrombeingoutput.
Thisfunctionhelpsreduceerroneousoutputwhentheowrateis“0”.
However, the unit of the low-cut value is the unit set in Subsection 5.1.4. If the unit is changed,
the low-cut value is changed to the converted value in conjunction with the unit.
Setthelow-cutvalueto“0”ifthelow-cutfunctionisnotused.
Thesettingcanbeconguredwiththefollowingparameters.
AI FB output
Menu path
Display Devicesetup►Detailedsetup►Fieldbusinfo►AI1FB*►(seebelow)
F
OUNDATION
Fieldbus
DeviceConguration►AI1*►DeviceConguration►Conguration►Options►(see
below)
*For AXG, one from Al1 to Al4, and for AXW, one from AI1 to AI3.
Block
Name
Relative
Index
Parameter
Description
Display F
OUNDATION Fieldbus
AI1 - AI4
(AXG)
AI1 - AI3
(AXW)
17 LOW_CUT LowCuto
Speciestheoutputlow-cut
value.
Asforlow-cutsettingsoftheoutputsignal(LOW_CUT),theoutputismadeto0iftheoutput
goes below the low-cut setpoint.
Toenablethelow-cutfunction,the“Lowcuto”optionshouldbeenabledwithIO_OPTS.
However, when the status returns from the low-cut state to the normal state, hysteresis is waited.
When the L_TYPE setting of the output mode is direct, the hysteresis width is 1% of XD_SCALE,
and otherwise, the width is 1% of OUT_SCALE.
F0448.ai
0
Output [m
3
/h]
Hysteresis
1 %
Input [m
3
/h]
low-cut value