6166 DC Voltage Current Source Operation Manual
9. SPECIFICATIONS
9-4
Thermal electromotive force source overall accuracy: At temperature of 23C5C and relative humidity
of 70 % or less for one year
Thermo-
couple
Source temperature - Relative
junction temperature
Accuracy
Range ( of setting + C)
T (CC) -220.0C to +400.0C
-220.0C to -190.1C 0.012 + 0.2C
-190.0C to -70.1C 0.009 + 0.2C
-70.0C to 50.0C 0.006 + 0.1C
50.1C to +400.0C 0.005 + 0.1C
J (IC) -210.0C to +1200.0C
-210.0C to -170.1C 0.006 + 0.1C
-170.0C to -100.1C 0.008 + 0.1C
-100.0C to +1200.0C 0.011 + 0.2C
E (CRC) -220.0C to +1000.0C
-220.0C to -190.1C 0.012 + 0.2C
-190.0C to -80.1C 0.009 + 0.1C
-80.0
C to 1000.0C 0.006 + 0.1C
K (CA) -220.0C to +1372.0C
-220.0C to -190.1C 0.012 + 0.3C
-190.0C to -130.1C 0.010 + 0.2C
-130.0C to -80.1C 0.006 + 0.1C
-80.0C to +1240.0C 0.006 + 0.1C
+1240.1C to +1372.0C 0.007 + 0.1C
S (PR10) -10.0C to +1768.0C
-10.0C to +50.0C 0.006 + 0.5C
+50.1C to +200.0C 0.006 + 0.4C
+200.1C to +1768.0C 0.006 + 0.3C
R (PR13) -10.0C to +1768.0C
-10.0C to +40.0C 0.006 + 0.5C
+40.1C to +160.0C 0.006 + 0.4C
+160.1C to +1768.0C 0.006 + 0.3C
B (PR30) +280.0C to +1820.0
C
+280.0C to +500.0C 0.004 + 0.9C
+500.1C to +650.0C 0.004 + 0.7C
+650.1C to +950.0C 0.004 + 0.5C
+950.1C to +1550.0C 0.004 + 0.4C
+1550.1C to +1820.0C 0.004 + 0.3C
N -220.0C to +1300.0C
-220.0C to -210.1C 0.015 + 0.4C
-210.0C to -180.1C 0.013 + 0.3C
-180.0C to -30.1C 0.009 + 0.2C
-30.0C to +1300.0C 0.006 + 0.1C