EasyManua.ls Logo

BERTHOLD TECHNOLOGIES Uni-Probe LB 490 - Page 164

BERTHOLD TECHNOLOGIES Uni-Probe LB 490
730 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38477BA2B
2 – 164 25.5.09
2 Installation Volume 2
Function failure due to damage
The detector fixture must not transfer any vibrations or heat onto
the shielding; otherwise the locking mechanism may be damaged
and the shielding effectiveness may be adversely affected.
Therefore, install the fixture on a vibration-free support or attenu-
ate possible vibrations using vibration absorbers. Prevent heat
transfer by using suitable insulating materials.
Install shielding 45°
IMPORTANT
Before installing the source shielding or the source, you have to
determine the background for rod detectors (see Volume 3, section
4.3.1 on page 3–351).
Figure 2-23 Point source with rod detector
The standard arrangement (Figure 2-23) shows the mounting posi-
tion of the shielding with point source; G indicates the topmost
point of the measuring range that can be covered. The shielding
must be mounted on a suitable fixture provided by the customer in
the appropriate height, so that the upper edge of the shielding is
positioned above the measuring range offset by “G”.
Size G for various shielding types
Shielding type Size G (in mm)
100 115
150 115
200 155
270 200
Uni-Probe LB 490
Point source
shielding
Mounting
bracket
Marking groove
G

Table of Contents