6 Technical Drawings Volume 2
6.3.5 Water Cooling with Horizontal Installation
With horizontal installation of the detector, the water inlet is con-
nected directly to the end of the pipeline on the side of the cable
entries.
6.3.6 Water Cooling with Vertical Installation
With vertical installation of the detector, the lowest nozzles have to
be used as water inlet. Otherwise, an air bubble may form which
may have a negative effect on water cooling.