80 | Chapter 2: Installation
Verify the Peri-Pump Dispenser
1. Fill the Peri-pump’s supply bottle with approximately one liter of deionized
water.
LHC
Touch screen
1. Select File > Open.
2. Open the MultiFlo FX and
Maintenance Protocols folders, and
then open P-xUL_CASS_RINSE.LHC*
You may need to reset the Com Port.
3. When ready, click the Run button to
prime the tubing and manifold with
deionized water.
4. When finished, close the program.
1. Select Quick at the Home screen.
2. Select Device Prime and set the
Device to the Peri-pump
Primary.
3. Press the PRIME button.
Repeat for Peri-pump Secondary, if
applicable.
4. When done, press the Home
button.
*Verify both a standard cassette and a RAD technology cassette, if applicable.
If leaks were detected, tighten all tube fittings and re-run the program. If leaks are
still detected, contact BioTek’s Technical Assistance Center.
Verify the Syringe Dispensers
1. Fill the supply bottles with approximately one liter of deionized water, one for
each syringe pump.
LHC
Touch screen
1. Select File > Open.
2. Open the MultiFlo FX and
Maintenance Protocols folders, and
then open S-DAY_RINSE_A&B.LHC
You may need to reset the Com
Port.
3. When ready, click the Run button
to prime the tubing and manifold
with deionized water.
4. When finished, close the program.
1. Select Quick at the Home screen.
2. Select Device Prime and set the
Device to the Syringe Both.
3. Press the Start button.
4. When done, press the Home
button.
MultiFlo™ FX Multi-Mode Dispenser