Verify Performance | 79
Verify Performance
Before using the MultiFlo FX for the first time, verify that it is operating properly.
l When using the LHC, make sure the MultiFlo FX is connected to the PC and both are
powered up.
l When running standalone, turn on the MultiFlo FX.
Using the touch screen:
Press Instrument>About >Self-Check.
Using the LHC:
1. Click the Name link on the main page and, if required, select the MultiFlo FX.
2. Define the COM Port used to connect the MultiFlo FX to the computer and
Test Communication.
3. In the Target Instrument Settings dialog that opens, click Get actual settings
now, and click OK.
4. Select Tools>Instrument Utilities
5. On the General Settings tab, click the Perform Self-Check link.
Test results:
l Pass: no error message is displayed.
l Fail: an error message is displayed. If this happens, note the error code and refer to
Troubleshooting on page 272 to determine its cause. If the problem is something you can
fix, turn off the instrument, fix the problem, and then turn the instrument back on.
Otherwise, contact BioTek’s Technical Assistance Center.
n Note: An instrument qualification package (PN 1260521) for the MultiFlo
FX is available for purchase from BioTek. The package contains thorough
procedures for performing Installation Qualification, Operational Qualification
and Performance Qualification (IQ/OQ/PQ) and preventive maintenance (PM).
Extensive Checklists and Logbooks are included for recording results.
n Important! Before operating the instrument, review Optimize Performance
on page 98. The guidelines include necessary steps to perform before running
a protocol, and issues to consider when creating or editing protocols.