EasyManua.ls Logo

Biotek MultiFlo FX - Liquid Testing the MultiFlo FX Multi-Mode Dispenser

Biotek MultiFlo FX
330 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Peri-pump Dispenser Maintenance | 197
RAD8-to-1 Cassette - Unclog the Tips
Note: Reassembling the 8-to-1 cassette is somewhat challenging. Before resorting to this
method of unclogging the tips, try multiple purge-prime rounds and/or autoclaving the
cassette. If that fails, proceed with this method.
Disassemble the cassette to unclog the tips:
1. Use the Philips screwdriver provided
with the MultiFlo FX to remove the
three screws on the 8-to-1 cassette's
cover plate.
2. Remove the plate and set it aside.
3. Remove the two screws that hold the
chute on the cassette.
4. Remove the chute and set it aside. This
frees the tips and tubing from the
cassette.
5. Pull the tubing off the tips and follow the instructions to use the 10 cc syringe
to unclog tips. See Unclog the Dispense Tips on page 194.
6. When finished, reassemble the cassette:
l Observe the numbers next to the tubing posts on the back side of the
molded tip head. Use the Center Holder to determine the proper order for
BioTek Instruments, Inc.

Table of Contents

Related product manuals