176 | Chapter 4: Maintenance
Recommended Maintenance Schedule
Tasks
Frequency
Daily Overnight/
Multi-Day
Weekly Periodic/
Monthly
Before
storage/
shipment
Washer
Run W-DAY_RINSE
ü ü
Run AutoPrime
ü
Run W-OVERNIGHT_LOOP
ü
Run W-RINSE_AND_SOAK
ü
Peri-pump
Flush dispense cassette
ü ü
Record number of plates
processed with cassette
ü
Syringe Dispenser
Run S-DAY_RIN SE_A&B
ü ü
Run AutoPrime
ü
Components
Remove protein residuals and
fungi growth, (if necessary)
ü ü ü
Check/empty waste bottles
ü ü
Clean bottles
ü ü
Clean plate carrier system
ü ü
Clean washer manifold
ü ü
Clean aspirate and dispense tubes
ü ü
Clean exterior surfaces
ü
Clean Syringe dispenser manifold
ü ü
Clean Syringe dispenser tubes
ü ü