Wash Parameters Table | 115
Advanced Options
Parameters Description
Bottom Wash
Height:
Z-axis Position
The distance between the bottom of the aspirate
tubes and the carrier surface. The value in mm is
displayed on screen.
Horizontal X-
axis Position
The left and right position of the dispense tubes
when the carrier is beneath the manifold.
Horizontal Y-
axis Position
The front-to-back carrier position (or Y axis) to align
the microplate with the manifold tubes during a
dispense.
Delay start of
Vacuum
until dispensing this volume.
Change the Plate Type
Edit a protocol to change its plate type:
1. Select Protocols at the home screen.
2. Highlight the protocol and press Edit.
3. Press the Info button on the step definition screen and touch the Plate Type
field to change the plate type.
4. Press OK to save the change. Press Save at the step definition screen to save the
protocol.
Rename a protocol
There are two ways to rename a protocol, edit the name or make a copy and
rename the copy. The MultiFlo FX behaves like Windows' Save As function,
creating a copy of the protocol file when it is renamed.
To change the protocol name:
1. Select Protocols at the Home screen.
2. Highlight the protocol and press Edit.
3. Press Info and touch the Name field.
4. Press OK and Save the protocol when you're done.
You can delete the originally named protocol to remove it from the Home screen.