218 | Chapter 5: Qualification
System Self-Test, Verify Information
Perform these steps to verify software information and run a system self-check:
Prerequisite: When controlling the instrument with the LHC, ensure that it is
attached to the host computer and turned on, and then launch the LHC software.
To run the System Self-Test:
LHC Touch screen
Select Tools>Instrument
Utilities>General Settings.
Under Instrument Functions,
click Perform Self-check.
1. Select Instrument at the Home screen.
2. On the Other screen, press Run Self-
Check.
Test Results:
l If the test fails, an error code displays. If this happens, find the error code in
the MultiFlo FX Operator’s Manual to determine its cause. If the problem is
something you can fix, turn off the instrument, fix the problem, and then turn
the instrument back on and re-run the test. If the problem is not something
you can fix, contact BioTek’s Technical Assistance Center.
Record Onboard Software
Record the software part number and version installed on the MultiFlo FX.
LHC Touch screen
1. Select Tools>Instrument
Utilities
2. Select the Software tab and from
the Basecode Software
Information section record the
l
Software Version and part
number
1. Select Instrument at the Home
screen.
2. Record the Software part number
and version from the About screen.
MultiFlo™ FX Multi-Mode Dispenser