58 | Chapter 2: Installation
Install RAD Technology Cassette
RAD technology's single-tube and 8-to-1 bulk dispensing cassettes are installed like
standard cassettes with one difference, the tip holder is installed on the dedicated
RAD dispense head. Install the RADcassettes on the external or secondary Peri-
pump only! They are not designed for use on an internal Peri-pump.
Follow instructions to Install the Dispense Cassette on page 56 skipping the first step.
Magnets on the dispense head hold the cassette tip(s) in place:
Single tube RADcassette 8-tube-to-1 Bulk RADcassette
Align the two pins with the holes (above and below the magnets) on the dispense
head.
n You must create a protocol to successfully use the RAD8-to-1 bulk-dispensing
cassette. It is not compatible with Quick Dispense.