EasyManua.ls Logo

Biotek MultiFlo FX - Predefined Sample Protocols

Biotek MultiFlo FX
330 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
122 | Chapter 3: Operation
Periodic Maintenance
S-Decontaminate Includes prompts for first running disinfectant and later running water
through the system. It can be easily modified to suit any major
cleaning effort. The A or B suffix identifies the syringe the protocol is
designed for.
W-LONG_
SHUTDOWN
Helps implement the recommended procedure for preparing the
instrument for storage. This protocol includes prompts for running
disinfectant, then water, and lastly, air through the system.
S-LONG_
SHUTDOWN
Prepares the instrument for long-term storage. This protocol includes
prompts for running disinfectant, then water, and lastly, air through
the system. Defined to use Syringe A or B.
W-PRIME_40 Simple prime routine.
QC (Quality Control) Protocols
Manifold-Specific
QC_96_DISP_TEST Dispense precision test protocol for 96-well wash manifold.
QC_96_EVAC_TEST Evacuation efficiency test protocol for 96-well wash manifold.
QC_SA-1536_
DISP_TEST
Dispense precision test protocol for 32-tube Syringe A manifold.
QC_SB-1536_
DISP_TEST
Dispense precision test protocol for 32-tube Syringe B manifold.
QC_P-1536_DISP_
TEST
Dispense precision test protocol for Peri-pump 1 µL cassette and
1536-well plate. (P2 protocols are for the secondary Peri-pump)
Predefined Sample Protocols
The "Sample" protocols are provided to facilitate learning. Some samples are model
specific. "P2" protocols are copies of the primary protocol but defined for the
Secondary Peri-pump.
LHC users: You may need to customize the protocols (as described on page 120) to
match your instrument's settings.
Serial dilutions
P-96_
DILUTION
Peri-pump serial dilution protocol dispenses 20 µL to 240 µL in 20 µL
increments to each column of the plate. 20 µL to column 1; 40 µL to
column 2; 60 µL to column 3; and so on till 240 µL to column 12.
MultiFlo FX Multi-Mode Dispenser

Table of Contents