294 | Appendix A: Error Codes
MultiFlo FX Software Error Codes
Generally, these errors are caused by protocol parameters that conflict with the
instrument's onboard settings. The protocol may have been originally created for a
different hardware configuration, the 192-tube wash manifold instead of the 96-
tube, for example.
Error
Code
Description Help
6000 General communication error during
download.
See Communication Errors
6001 COM port created by USB converter
no longer active.
See Communication Port
6002 Invalid basecode part number;
instrument is not MultiFlo FX.
Service Only.
Contact BioTek TAC.
6003 Invalid Basecode Data Version;
basecode needs to be updated.
Contact BioTek to obtain latest
basecode.
6004 No rows are selected for the specified
plate type.
Modify the protocol to select a row.
6005 Invalid row selection value (must be
0 or 1).
Contact BioTek TAC.
6006 This instrument can only process 96-
well plates.
The protocol may have been created for
another instrument, change the plate
type or select another protocol.
6007 This instrument can only process
1536-well plates.
6008 The 8-tube Syringe Manifold can only
be used with 96-well plates.
Mismatch between installed hardware
and protocol parameters: change the
plate type or correct the instrument
settings to match the currently installed
hardware.
6009 The 96-tube single wash manifold can
only be used with 96-well plates.
6010 The data is invalid or out-of-range. Service Only.
Contact BioTek TAC.
6011 This step type cannot be
downloaded.
Review the limitations to transferring
protocols to the instrument, See the
LHC Help Topic: Transferring Protocols.