Verify Performance | 81
If leaks were detected, tighten all tube fittings and re-run the program. If leaks are
still detected, contact BioTek’s Technical Assistance Center.
Verify the Washer Component
1. Fill the washer’s supply bottle with approximately one liter of deionized water.
LHC Touch screen
1. Select File > Open.
2. Open the MultiFlo FX and
Maintenance Protocols folders and
then open W-DAY_RINSE.LHC
You may need to reset the Com
Port.
3. When ready, click the Run button
to prime the tubing and manifold
with deionized water.
4. When finished, close the program.
1. Select Maintenance at the Home
screen.
2. Select W-Day_Rinse and press
Start.
3. When the protocol is completed,
press the Home button.
If leaks were detected, tighten all tube fittings and re-run the program. If leaks are
still detected, contact BioTek’s Technical Assistance Center.