140 | Chapter 3: Operation
When an 8-channel dispenser addresses a 1536-well plate, it first dispenses to rows
1, 5, 9, 13, 17, 21, and 25. The next section of rows dispensed to includes 2, 6, 10, 14,
18, 22, 26, and 30. And so on, creating 4 sections of rows.
Dispense Pattern with Both Syringes
When "Both" Syringe dispensers are selected to fill a plate, the manifolds move in
this pattern: