Dispense Processing Patterns | 139
8-Channel Dispenser Dispense Pattern
When addressing high density plates, 384- and 1536-well, the 8-channel
cassette/syringe dispenser moves in the pattern described here to fill the plate.
With Advanced Dispense Options, you can skip certain rows and columns.
The MultiFlo FX let's you specify which columns to skip during a dispense step.
384-Well Plate
Rows: You can skip one of the two "rows" sections:
When an 8-channel dispenser addresses a 384-well plate, it first dispenses to odd
numbered rows, and then dispenses to even numbered rows. With this feature you
can dispense to odd rows in one dispense step, change fluids or use another
dispenser and dispense to even rows in another dispense step.
In addition, you can combine a selection of columns with one of the two row
sections (odd or even) to define a complex distribution pattern.
1536-Well Plate
Rows: You can skip up to three "rows" sections: