Peri-pump Advanced Settings | 159
Peri-pump Advanced Settings
Settings for 0.5 µL Dispense
Do not alter these settings without explicit instructions from BioTek.
Peri-pump offsets are required to ensure precise and repeatable dispense
performance when dispensing 0.5 µL. The default values were determined after
thorough testing and should not be changed. This control is provided in
anticipation of potential future changes to the 1 µL cassette tubing. In that case,
explicit instructions to alter the offsets will be provided.
Contact BioTek TAC for assistance if you have questions about dispense performance
at 0.5 µL.
8-to-1 Tube RAD Cassette
Chute height represents the distance between the bottom of the 8 tube tips to the
bottom of the chute tip. The default value reflects the chute provided with the
bulk-dispensing cassette. Only change this value if you remove or replace the
chute.
Set the Chute Height this way:
l If you remove the chute and run the cassette without one,
enter 0.
l If you replace the chute, measure from the bottom of the 8
tube tips to the bottom of the replacement chute, and
enter that value in millimeters (mm).
Contact BioTek TAC for assistance.