MultiFlo FX Software Error Codes | 295
Error
Code
Description Help
6012 Illegal characters in protocol name. See the LHC Help topic: Define a
Protocol.
6013 The protocol name length must be
16 characters or less.
6014 A 1536 well plate is not supported
by this instrument.
Service Only.
Contact BioTek TAC.
6015 The specified volume exceeds the
cassette maximum limit.
Modify the volume or change the
cassette type.
6016 The volume is out-of-range. Modify the volume or change the
cassette type.
6017 Invalid flow rate. Learn about the Syringe Dispense Step
6018 Invalid number of
pre-dispenses.
Service Only.
Contact BioTek TAC.
These codes indicate an unexpected
software error that cannot be fixed
without BioTek support.
6019 Invalid horizontal dispense position.
6020 Invalid dispense height.
6021 Invalid plate clear height.
6022 Invalid column selection value
(must be 0 or 1).
6023 Invalid protocol step type.
6024 The Definition String contains
invalid data.
6025 Manifold conflict between protocol
requirements and instrument
configuration.
Change the Washer Manifold or change
the Instrument Setting.
6026 Valve module conflict between
protocol requirements and
instrument configuration.
Make sure the Buffer Switching setting
matches your instrument; see
Instrument Settings.
6027 Syringe module conflict between
protocol requirements and
instrument configuration.
Make sure the Syringe dispenser setting
matches your instrument: see
Instrument Settings.
6028 Filter washer conflict between
protocol requirements and
instrument configuration.
Service Only.
Contact BioTek TAC.
6029 Required cassette does not match Change the cassette type to match the
BioTek Instruments, Inc.