296 | Appendix A: Error Codes
Error
Code
Description Help
installed cassette.
Using RAD technology cassettes
increases the chance of getting this
error. Be as precise as possible
when using this feature.
protocol requirement and rerun the
protocol.
6030 Invalid cassette type was specified. Service Only.
Contact BioTek TAC.
6031 Cannot use a 96-well plate with a
192-tube manifold.
Modify the Plate Type or Change the
Washer Manifold.
6032 Downloading Protocols is not
supported.
Service Only.
Contact BioTek TAC.
6033 This step is not supported for 1536-
well plates.
Fix the plate type or the Instrument
Settings. A conflict between the plate
type and installed hardware devices has
been detected.
Change the Plate Type to one supported
by the washer/dispenser. Or, click the
Instrument Settings link and make sure
they match the physical hardware: Get
settings from instrument.
6034 The 32-tube Syringe Manifold is
required for 1536-well plates.
6035 The 16-tube Syringe Manifold is
required for 96- and 384-well plates.
6036 The 128-tube Washer Manifold is
required for 1536-well plates.
6037 The 128-tube Washer Manifold can
only be used for 1536-well plates.
6038 This step only applies to 1536-well
plates.
6039 Conflicting column selection. Fix the plate map (selected columns to
dispense to). A protocol parameter may
have been changed after a partial plate
dispense was defined.
6040 Invalid baud rate Service Only.
Contact BioTek TAC.
These codes indicate an unexpected
software error that cannot be fixed
without BioTek support.
6041 Invalid data bits selection
6042 Invalid stop bits selection
6043 Invalid parity selection