MultiFlo FX Software Error Codes | 297
Error
Code
Description Help
6044 Serial port error Fix the COM port setting. Check the
cabling. Click the Port link and use the
drop-down menu to see all active ports.
Customize the Predefined Protocols to
avoid this error in future.
When controlling the BioStack with the
LHC, make sure the instrument's
BioStack setting is Manual.
6045 Serial write error
6046 Serial read error
6047 Checksum error Contact BioTek TAC.
6048 Serial NAK error Make sure the COM port setting is correct
and the cable is properly connected.
Restart the instrument. If error reoccurs,
contact BioTek TAC.
6049 Excess data, or not enough data,
received.
To correct these errors:
l Reset the instrument.
l Check cables, plug in only one
communication cable at a time:
USBor serial.
l Try running a different protocol.
If error reoccurs, contact BioTek TAC.
6050 Invalid message header
6051 Invalid message object
6052 Invalid message body size
6053 Serial message timeout
6054 Port handle error
6055 Read timeout value is invalid.
6056 Unauthorized to open the COM port. Make sure the COM port setting is correct
and the cable is properly connected.
Restart the instrument. If error reoccurs,
contact BioTek TAC.
6057 Out-of-range parameter for the open
port function.
6058 Unable to open the COM port.
6059 Unable to clear the transmission
buffer.
6060 Unable to close the port.
6061 Port is no longer available.
6062 Unhandled exception while
transmitting message.
Contact BioTek TAC
6063 The selected plate type is not allowed
with this protocol step.
Modify the protocol to change the plate
type.