Peri-pump Dispense Precision & Accuracy Test Worksheet
8-to-1 RAD Cassette Test
100 µL Dispense Precision Test
Standard Deviation (SD):
Mean Absorbance (sum of all wells ÷ 24)
% CV (SD ÷ Mean x 100) %
% CV must be < 2.5% c Passc Fail
100 µL Dispense Accuracy Test
Total Dispense (Actual) Weight: grams
Expected Weight:
(mL/well x number of wells dispensed)
grams
% Accuracy Error:
(Actual Weight Expected Weight) ÷ Expected Weight x 100
%
% Accuracy Error must be < 4.0% c Passc Fail
Visual verification that no well varies
considerably from the others
c Passc Fail
Cassette Serial Num-
ber:
Tests Performed By:
Date:
Reviewed/Approved
By:
Date:
MultiFlo FX Qualification Procedures