Evacuation Efficiency Test Worksheet
96/384-Well Manifold
96-Well Plate
Serial Number:
Standard Test
Total Residual Weight: g
Verification that wells are consistent in appearance: q Pass q Fail
Mean Residual Weight (Total Residual Weight ÷ 96): g
Mean Residual Weight <= 0.002 g? q Pass q Fail
Evacuation Diagnostics Test (check here q if not performed)
Mean OD for the plate (Sum of all wells ÷ 96):
Residual Factor (Mean OD ÷ Mean Residual Weight):
Calculate the Residual Weight for each well: well OD ÷ Residual Factor
Every Residual Weight per well <= 0.002 g? q Yes q No
Date:
Test Performed By:
MultiFlo FX Qualification Procedures
© BioTe Instruments, Inc.