24 | Chapter 1: Introduction
Dispense Precision
Refer to the values below for washer dispense precision when measured in full plates with
the specified volume of deionized water with 0.1% Tween 20 and FD&C#1 blue dye
solution and read at 630 nm/405 nm.
Plate Type Performance Volume L/well)
96-Well ≤3.0% CV 300
6-Well ≤5.0% CV 5560
12-Well ≤3.0% CV 2240
24-Well ≤3.0% CV 1120
48-Well ≤3.0% CV 560
384-Well ≤3.0% CV 80
Dispense Accuracy
Washer dispense accuracy shall be ≤±3.0% when deionized water with 0.1% Tween 20
(and FD&C#1 blue dye) solution. The weight of the fluid dispensed shall be measured
gravimetrically.
Plate Type Volume L/well)
96-Well 300
6-Well 1120
12-Well 2240
24-Well 1120
48-Well 560
384-Well 80
MultiFlo FX Multi-Mode Dispenser