EasyManua.ls Logo

GE Voluson E Series - Save Small Backup (Scan Settings)

GE Voluson E Series
356 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.4.1 Save Small Backup (Scan Settings)
The Image/Scan Settings contain:
2D Factory and 2D User Presets
3D/4D Factory and 3D/4D User Presets
Annotation Presets
Scan Assistant Configuration
Measure Configuration
Biopsy Lines
1.
Insert a DVD/CD+R/RW into the drive or connect an external USB device.
2. Press the Utilities key on the control console.
3. In the “Utilities” menu touch the
Setup
button to invoke the setup desktop on the screen.
4. On the right side of the screen select
Backup
and then click the
System Configuration
tab.
5. Click the
Save
button (1) of the "Small Backup (Scan Settings)" group.
Figure 4-10 Save window
6. Choose the media (2) and click the
Save
button (3).
7. Select the
New File...
key and enter a file name (without extension).
8. Click the
OK
key to start the process. When the saving has been completed, click
OK
.
Functional Checks
Voluson E-Series Service Manual
5539550APB Revision 6
4-11

Table of Contents